Addendum describing the Respiration Rate Version 1.0 parameter for the Nellcor bedside respiratory patient monitoring system.
This addendum provides detailed information on the Respiration Rate Version 1.0 parameter, including its operation, safety considerations, and performance specifications. Caregivers can use the parameter alongside SpO₂ and pulse rate monitoring without changing existing system settings.
The document also covers installation guidance, data output fields, troubleshooting tips, and clinical study results that validate the accuracy and precision of the respiration rate measurement.
| Respiration Rate accuracy | ±1 breath per minute |
|---|---|
| Upper alarm threshold | 28 breaths per minute |
| Lower alarm threshold | 8 breaths per minute |
| Pulse rate range | 20 to 250 beats per minute |
| SpO₂ saturation range | 1% to 100% |
The measurement is accurate to within ±1 breath per minute compared to a capnography‑based reference.
Alarms are set at a high threshold of 28 breaths per minute and a low threshold of 8 breaths per minute.
No, the Respiration Rate parameter is not intended for use as an apnea monitor.