Nellcor OXIMAX N-65 Operator's Manual
The Nellcor OXIMAX N-65 is a prescription device intended for attended monitoring of oxygen saturation (SpO2) and pulse rate in adult and neonate patients. It features an automatic calibration system, audible indicators, and the ability to record sensor event data. The device operates using Nellcor OXIMAX sensors and includes options for battery operation or external power.
Read onlineNellcor N-65 specifications
| Peak Wavelengths | Red: 660 nm, Infrared: 900 nm |
|---|---|
| Maximum Output Power | Less than 15 mW |
| Default SpO2 Alarm Limit (Neonatal Mode) | 80% |
| Data Update Period (Normal Conditions) | 6 to 7 seconds (3 seconds in Fast Mode) |
Source: Nellcor N-65 Pulse Oximeter Addendum to Monitor Instructions
All Nellcor N-65 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the maximum output power of the N-65 sensor LEDs?
The total optical output power is less than 15 mW.
From: Nellcor N-65 Pulse Oximeter Addendum to Monitor Instructions
What happens if the dynamic averaging time for SpO2 exceeds 20 seconds?
The PULSE SEARCH indicator illuminates solid, and SpO2 and pulse rate update every second.
From: Nellcor N-65 Pulse Oximeter Addendum to Monitor Instructions
Is the OXIMAX N-65 defibrillator-proof?
No, the device is not defibrillator-proof. Measurements may be inaccurate during or shortly after defibrillation.
Who should operate the OXIMAX N-65?