The GMA 313 is a CO2 gas monitor produced by GfG. It can be used to detect the concentration of CO2 gas in the working environment.


GfG GMA 313 CO2 Gas Monitor Operation Manual - cover page
Brand
GFG
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
1 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
E15428DE8379C30707CB5730A7C85A61

About this manual

The GMA 313 is a compact CO2 gas monitor using infrared detection for continuous ambient air monitoring. It provides audible and visual alarms when CO2 levels exceed preset thresholds, with a relay for external control. The device requires a 30-minute warm-up before operation and includes service functions for adjusting alarm thresholds using a calibration adapter. Maintenance should be performed annually or as needed to ensure accurate readings and safety compliance.

Key features include IP54 protection, a diffusion inlet, and optional mounting brackets. The unit operates on 110V AC, with alarm thresholds set at 1.5% and 3.0% CO2. Faults are indicated by red and yellow LEDs, and the device can be reset locally or remotely via the optional GMA 313 EQ unit.

What's inside

  • Introduction
  • For Your Safety
  • Application and Purpose
  • Detection Principle - Infra-red
  • Operation Notes
  • Turning On
  • Function Test
  • Detection Mode
  • Display and Operation Elements
  • Alarm for Gas Hazards

Specifications

Detection PrincipleInfra-red (NDIR)
Relay OutputMax. 250 V AC, 5 A

Frequently asked questions

How long does the GMA 313 need to warm up before operation?

The GMA 313 requires a 30-minute warm-up period before it can detect CO2 concentrations.

How do I reset the alarm?

Press the reset button “Q” to reset the alarm. The buzzer can be reset at any time by pressing “Q”.

What is the protection rating of the GMA 313?

The GMA 313 has an IP54 protection rating against water and dust.