GfG GMA44 Gas Detection System Operation Manual

This document is the GMA44 operation manual, which introduces how to use this product, including precautions before use, detection mode, display screen, alarm configuration, fault troubleshooting, etc.


GfG GMA44 Gas Detection System Operation Manual - cover page
Brand
GFG
Category
Industrial Supply
Document type
User Manual
Language
English
Pages
25
File format
PDF
File size
1.1 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
D183C0B9877832A5F65A0B8441E0E300

About this manual

The GfG GMA44 is a fixed gas monitoring controller designed to connect up to four transmitters of the same type. It monitors ambient air for toxic, combustible, or oxygen levels and provides a 3-digit display for measurement values. The system supports various detection modes and allows for adjustable alarm thresholds and relay functions to suit specific safety requirements.

The manual details the operation of the device, including how to turn it on, perform self-tests, and configure alarm parameters. It also covers the service mode for adjusting zero points and sensitivity, as well as maintenance procedures such as regular function tests and calibration instructions.

What's inside

  • Introduction
  • Application
  • Distinguishing the Types of GMA44
  • For your Safety
  • Detection Mode
  • Function Description
  • Turning On
  • Alarm configuration
  • Fault
  • Relays

Specifications

Max. number of channels4
Supply voltage (GMA44)24 V DC
Supply voltage (GMA44 EC)230 V AC / 24 V DC
Number of relays4
Number of alarm thresholds3

Frequently asked questions

How many transmitters can the GMA44 connect?

The GMA44 allows to connect up to four transmitters of the same type.

What indicates a fault in the GMA44?

A fault is indicated when the yellow LED 'SF' lights up and the fault relay is activated.

How do I acknowledge an activated alarm?

Press the key 'QUIT MENU' to acknowledge the alarm.