GfG EC 21 Fixed Transmitter Operations Manual

This manual describes the basic operation of the GfG EC 21 transmitter. It covers the transmitter's features and operation, as well as how to perform zero point adjustment, sensitivity check, and percent of scale to mV output.


GfG EC 21 Fixed Transmitter Operations Manual - cover page
Brand
GFG
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
248 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
A24E723826C87692E3C35AA950FFB2BF

About this manual

This manual provides instructions for operating and maintaining the GfG EC 21 fixed transmitter, including safety precautions, electrical connections, zero point adjustment, sensitivity calibration, and troubleshooting. It details how to connect power, adjust zero and span using test terminals, and interpret mV output for gas concentration. Technical specifications such as voltage, output signal, and environmental limits are included.

The document also covers common issues like no output or sensor failure and provides solutions, including checking wiring polarity, replacing expired sensors, and verifying calibration gas concentration.

What's inside

  • For Your Safety
  • General Description
  • Electrical Connections
  • Zero Point Adjustment
  • Sensitivity Check (Calibration)
  • Percent of Scale to mV Output
  • Technical Data
  • Troubleshooting

Specifications

Output Signal4–20mA Linear
Dimensions2.625 x 5.25 x 6 in.
Temperature Range-20 to 50° C
Relative Humidity15–95% rh non-condensing

Frequently asked questions

What is the output voltage for 100% scale on the EC 21?

The output voltage is 200 mV, corresponding to 20 mA.

What should I do if the transmitter has no output?

Check if the supply wires are reversed; correct polarity is required for operation.