This guide provides helpful instructions, tips and troubleshooting suggestions for calibrating a ProDSS instrument. For more detailed information on calibration and information on how to setup and operate a ProDSS, please refer to the ProDSS User Manual.
This document serves as a practical guide for calibrating the YSI ProDSS instrument, offering helpful instructions and troubleshooting advice for various sensors. It emphasizes the importance of using fresh, traceable calibration solutions and following standard operating procedures for accurate results. The guide includes a calibration worksheet to document performance and track sensor health.
| Acceptable Conductivity Cell Constant Range (626902 sensor) | 4.5 to 6.5 |
|---|---|
| Acceptable Conductivity Cell Constant Range (Integral sensors) | 4.4 to 6.4 |
| Acceptable ORP Cal Offset Range | -100 to 50 |
| Acceptable pH Slope Range | ~ 55 to 60 pH/mV |
| Ideal pH Slope | 59.16 mV/pH |
| Acceptable DO ODO Gain Range | 0.75 to 1.50 |
The conductivity calibration should be verified every day the instrument is used, although the sensor is stable and may hold calibration for several weeks.
The shelf life of conductivity solution is one month after being opened due to potential changes caused by evaporation.
Use 1.0 mS/cm for fresh water, 10 mS/cm for brackish to estuarine water and 50 mS/cm for salt water.