Fluke 1520 MegOhmMeter Calibration Manual

This document is the warranty information for Fluke 1520 MegOhmMeter Calibration Manual


Fluke 1520 MegOhmMeter Calibration Manual - cover page
Brand
Fluke
Document type
Service Manual
Language
English
Pages
24
File format
PDF
File size
272 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
1655CF14D32AA893D407AD34AB3C715C

About this manual

This manual provides calibration and maintenance instructions for the Fluke 1520 MegOhmMeter. It covers safety precautions, environmental and electrical specifications, and procedures for testing and replacing batteries and fuses. The document also includes performance test methods for verifying accuracy in volts, ohms, and megohms.

The manual is intended for qualified personnel only and references the Fluke 1520 User Manual for full operating instructions. It includes tables and figures for international symbols, required equipment, and fuse replacement steps.

What's inside

  • Introduction
  • Contacting Fluke
  • Precautions and Safety Information
  • International Symbols
  • Specifications
  • Required Equipment
  • Performance Test
  • Display Test
  • Backlight Test
  • Battery Test

Specifications

Operating Temperature-10 °C to +50 °C
Storage Temperature-40 °C to +70 °C
Battery TypeC Size 1.5 V alkaline, ANSI/NEDA-14A, IEC-LR14 (4 pieces)
Fuse6 mm x 32 mm (0.25 x 1.25 inch), 0.5 A 660 V, Fast Acting, 50 A Minimum Interrupt rating
Insulation Resistance Display Ranges4.000 M Ω, 40.00 M Ω, 400.0 M Ω, 4000 M Ω
Accuracy (MegOhm)10 % + 2 counts + 1 %/1000 M Ω
Test Voltages250 V, 500 V, 1000 V
Size23.4 x 10 x 6.4 cm (9.2 x 3.9 x 2.5 in)
Weight1 kg (2.2 lbs.)

Frequently asked questions

What is the warranty period for the Fluke 1520 MegOhmMeter?

The warranty period is three years and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days.

How do I test the internal fuse of the Meter?

Turn the rotary switch to the Lo Ω Function. Connect the test leads to the input terminals and short them together. Press TEST. The display should indicate approximately 0.5 Ω. If the display reads >40 Ω, replace the fuse.

What is the maximum operating voltage?

Maximum Operating Voltage 600 V ac or dc between any terminal and earth ground.