User's Guide ExStik® EC500 pH/Conductivity / TDS / Salinity / Temperature Meter
The ExStik EC500 is a handheld meter designed to measure pH, Conductivity, Total Dissolved Solids (TDS), and Salinity using a single electrode with dynamic cell-constant technology. The device operates on four CR2032 Lithium Ion Batteries and features an LCD display with indicators for battery status, measurement mode, and calibration reminders.
| Power Source | four (4) CR2032 Lithium Ion Batteries |
|---|---|
| Auto Power Off | 10 minutes of inactivity |
| TDS Compensation Ratio Range | 0.4 to 1.0 |
| Salinity Mode Ratio | 0.5 |
| Max Stored Readings | 25 |
| pH Calibration Points | 1, 2, or 3 points |
| Renew Indicator Threshold | 70% of a nominal slope |
Press and hold the MODE/HOLD button for 2 seconds until the display scrolls through the units (Conductivity, pH, TDS, Salinity), then release when the desired unit is displayed.
A flashing 'RENEW' warning indicates that the probe is not performing to expected specifications, likely because the pH electrode slope has fallen below 70% of a nominal slope.
Press the MODE/HOLD button to store a reading. The storage location number will appear, and the 'HOLD' indicator will be displayed.
With the unit on, press and hold the ON/OFF button for 4 seconds until 'clr' is briefly displayed.