This is a user manual for the wp 600 series meters, which includes instructions for the use of various functions of the meter and the setup and calibration methods of the meter
This manual provides step-by-step instructions for setting up, calibrating, and using Thermo Fisher Scientific's WP 600 Series Meters. It covers all models including pH 600, 610, 620, COND 600, DO 600, and others. The guide includes sections on general setup, temperature calibration, specific parameter calibration (pH, conductivity, dissolved oxygen, etc.), measurement modes, data logging, and transferring data via IrDA or USB. It also details technical specifications and accessories for each model.
The manual emphasizes the importance of completing the general setup and calibration steps before using specific parameter guides to ensure accurate measurements. It includes detailed procedures for calibration with standard or user-defined buffers, automatic temperature compensation, alarm settings, and data transfer to computers using CyberComm 600 software.
| Temperature (pH and Conductivity) Range | -10.0 ºC to 110.0 ºC (14.0 ºF to 230.0 ºF) |
|---|---|
| mV Range | ± 2000.0mV |
| Ion Range | 0.001 to 19900 Units ppm, molar, mg/L |
The meter has Automatic Temperature Compensation (ATC) which adjusts measurements based on temperature. Manual Temperature Compensation (MTC) is also available.
This manual covers all WP 600 Series Meters including pH 600, 610, 620, COND 600, 610, DO 600, PC 650, PD 650, CD 650, PCD 650.