This manual introduces the operator's overview and maintenance guide for the Waters Xevo TQD mass spectrometer.
This document serves as an Operator's Overview and Maintenance Guide for the Waters Xevo TQD Mass Spectrometer. It outlines safety considerations, including hazards related to solvent leakage, flammable solvents, and high temperatures, as well as guidelines for handling contaminated components during decontamination. The guide is intended for users and technicians operating or servicing the instrument.
You must renew the source O-rings at intervals not exceeding one year.
The nitrogen supply pressure must never fall below 400 kPa (4 bar, 58 psi) during analyses.
Equipment placed on top of the instrument must not exceed 15kg.