The Waters Gen-Pak FAX column is a 4.6 x 100 mm steel column containing a polymer-based high-performance anion-exchange packing with 2.5 µm particles. It is designed for high-resolution analysis and purification of various nucleic acid species, including DNA restriction fragments, PCR products, plasmids, and synthetic oligonucleotides, up to at least 5,000 base pairs.

What's inside

  • Introduction
  • Installation
  • General Usage Considerations
  • General Method Development Guidelines
  • Example Using Hae III Digested DNA
  • Troubleshooting

Specifications

Particle Size2.5 µm
Column Dimensions4.6 x 100 mm
Maximum Operating Pressure4,000 psi
pH Range1.5 to 12
Recommended Flow Rate0.75 mL/min
Recommended Temperature30°C
Recommended Buffer A25 mM Tris/Cl, 1 mM EDTA, pH 8.0
Recommended Buffer B25 mM Tris/Cl, 1 mM EDTA, 1.0 M NaCl, pH 8.0

Frequently asked questions

What is the maximum operating pressure for the Gen-Pak FAX column?

Do not exceed 4,000 psi operating pressure or about 1 mL/min at 25 °C.

What is the recommended pH range for the column?

Stay within a pH range of about 1.5 to 12, avoiding concentrated acids or bases.

How should the column be cleaned between nucleic acid injections?

Clean the column with 3 to 5 mL of 20 - 40% acetic acid.

What is the procedure for long-term storage?

Flush with water and methanol/water mixture, disconnect, plug ends, and store at 4 °C.