Waters 2695 HPLC System Operating Procedures

This document provides operating procedures for the Waters 2695 HPLC system, covering startup, preparation, data acquisition, processing, shutdown, and…


Waters 2695 HPLC System Operating Procedures - cover page
Brand
Waters
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
261 KB
Published
14 October, 2026
Updated
14 October, 2026
MD5 checksum
AB977D80E380776CBC340D8F7A6C8103

About this manual

The manual details the step-by-step process for starting up the Waters 2695 High-Performance Liquid Chromatography (HPLC) system. It instructs users to power on network equipment and computers before turning on instrument units, followed by launching the chromatography workstation after self-checks are complete.

The guide covers essential preparation steps including degassing mobile phases, purging the pump unit using Dry or Wet Prime functions, and flushing the injector syringe once system pressure stabilizes. It also outlines procedures for data acquisition via single or sequence injections, post-run data processing and printing, and proper shutdown protocols involving solvent flushing to protect the instrument. This manual is written in Chinese.

What's inside

  • 开机
  • 仪器的准备
  • 进样数据采集
  • 数据处理及打印
  • 关机
  • 特别注意

Frequently asked questions

What steps are required to prepare the instrument before starting the chromatography workstation?

You must degas the mobile phase through a 0.45 µm filter and ultrasonically, then purge the pump and injector lines to ensure they are filled with the mobile phase.

How should the system be flushed if the experiment used an organic phase mixed with buffer salt water?

You must first flush the system with 100% water for no more than 20 minutes, followed by a flush with 100% organic phase.

What specific cleaning procedure is required for the pump head and injector needle after the experiment?

Select Prime SealWash to clean the pump head for about 1 minute, then select Prime NdlWash to clean the injector needle 1-2 times.