Endevco IM7264D Piezoresistive Accelerometer Instruction Manual

Instruction manual for the Model 7264D piezoresistive accelerometer


Endevco IM7264D Piezoresistive Accelerometer Instruction Manual - cover page
Brand
Endevco
Category
Other Electronics
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
122 KB
Published
18 October, 2023
Updated
18 October, 2023
Model
IM7264D
MD5 checksum
AEC6F1C6AF85E8386D53B5F8BEC927B1

About this manual

The Endevco IM7264D is a low mass, miniature piezoresistive accelerometer. The document outlines the materials required for mounting, including specific Allen wrenches and screws supplied with the unit. It emphasizes precautions such as avoiding drops and preventing over-torque during installation to avoid damaging the unit.

The manual details a check-out procedure to verify proper operation upon receipt, including impedance tests, insulation resistance checks, and zero measurand output verification. It also provides step-by-step mounting instructions, specifying surface roughness requirements and torque values, along with guidelines for cable management and recalibration intervals.

What's inside

  • Materials Required
  • Special Precautions
  • Check-out
  • Mounting
  • Installation
  • Recalibration
  • Cleaning

Specifications

Typelow mass, miniature piezoresistive accelerometer
Mounting Torque2.6 in-lbf (0.29 N-m) recommended, 3.0 in-lbf (0.34 N-m) maximum
Insulation Resistance100 megohms minimum
Recommended Surface Roughness32 microinch rms
Calibration Interval6 to 12 month intervals

Frequently asked questions

What precautions should be observed when handling the accelerometer?

Avoid dropping the accelerometer or striking the unit against hard surfaces, and keep the protective sleeve on until the unit is installed.

What is the maximum recommended mounting torque?

The maximum recommended mounting torque is 3.0 in-lbf (0.34 N-m).

What is the minimum insulation resistance reading?

All insulation resistance readings should be 100 megohms minimum.

How often should the accelerometer be recalibrated?

Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals.