The product is a HART temperature transmitter that provides multiple interfaces and supports multiple protocols, with automatic self-calibration and fault detection
This is the user manual for the PR 7501V104-UK field-mounted HART temperature transmitter. It provides detailed instructions for installation, calibration, HART communication, and device operation. The device supports RTD, TC, and mV inputs with 4–20 mA analog output and HART 7 functionality. It is designed for hazardous areas and includes a local operator interface with optical buttons. Technical specifications include operating temperature ranges, accuracy, and EMC immunity.
The manual also covers accessories, mounting, wiring, display settings, and programming via the local interface or HART. It includes installation drawings for various certification standards and lists approvals such as ATEX, IECEx, and FM. The device supports multiple languages and has configurable backlight and error indication colors. This manual is written in Chinese.
| Input Signals | TC, RTD, Ohm, bipolar mV |
|---|---|
| Output | 4...20 mA analog PV output and HART |
| Display Resolution | 96 x 64 pixels |
| Operating Temperature (Silicone O-ring) | -40°C to +85°C |
| Operating Temperature (FKM O-ring) | -20°C to +85°C |
| Supply Voltage (Ex ia) | 10...30 VDC |
| HART Protocol Revisions | HART 7 and HART 5 |
| Accuracy (TC B1) | > 400°C |
| Accuracy (TC B4) | < 85°C |
| Protection Degree | IP54 / IP66 / IP68 / type 4X |
| Dimensions | Ø 110 mm |
| Weight | 1.3 kg |
| Wire Size | 0.13...1.5 mm² / AWG 26...16 stranded wire |
| Isolation Voltage | 1.5 kVAC / 50 VAC |
| Signal/Noise Ratio | > 60 dB |
| Response Time (Programmable) | 1...60 s |
| Long Term Stability | 0.1% of span / year |
| Sensor Current | Nom. 0.2 mA |
| Max. Offset on Input Signal | 50% of selected max. value |
| Load Resistance | ≤ (Vsupply - 10) / 0.023 [Ω] with backlight |
| HART Start-up Time | Max. 5 s |
| EMC Immunity | NAMUR NE 21, A criterion, burst … < ±1% of span |
| Certifications | ATEX, IECEx, FM, CSA, INMETRO, EAC |
It supports RTD, TC, Ohm, and bipolar mV inputs.
It is -40°C to +85°C.
It supports HART 7 and HART 5.
It is 0.1% of span per year.