This manual introduces the use of the Transcend Heated Humidifier, including setup, water filling, power supply, starting therapy, cleaning and maintenance.
This user guide provides instructions for setting up, operating, and maintaining the Somnetics Transcend Heated Humidifier. It includes steps for filling the water reservoir, connecting the device to a PAP machine, and cleaning procedures. Safety warnings are included regarding proper placement, water handling, and electrical safety. The guide also covers troubleshooting and technical specifications for the device.
The humidifier is designed for single-patient use with compatible Transcend PAP therapy devices to humidify air during sleep apnea therapy. It must be placed lower than the patient’s position to prevent water from entering the air tubing. The device requires a separate power source and must be drained before transport to avoid electrical hazards.
| Operating Conditions | Line power quality should be that of a typical commercial or hospital environment. |
|---|---|
| Electromagnetic Immunity | Compliance with IEC 60601 test level for electromagnetic environment. |
| Conducted RF | 3 Vrms 150 kHz to 80 MHz |
Remove the Water Reservoir, pull up the Fill Cap, fill with distilled water to the MAX line, replace the cap, and reinsert the reservoir.
Only use positive airway pressure devices, accessories, or parts recommended by Somnetics with the Humidifier.
The warranty information is provided in the 'Warranty' section on page 22.