This is a sleep apnea therapy user manual
This manual provides instructions for using the Somnetics EZEX Sleep Apnea Therapy device, including setup, operation, and safety precautions. It details contraindications, warnings, and maintenance procedures for the device and its components. The device delivers positive airway pressure for treating obstructive sleep apnea in adults weighing over 66 pounds (30 kg).
| Transcend EZEX weight | Less than 1.0 lbs (454 g) |
|---|---|
| Transcend EZEX dimensions | 6.1 in x 3.5 in x 2.8 in (15.4 cm x 8.9 cm x 7.0 cm) |
| Air outlet connector port dimensions | 19-mm diameter proprietary connector |
| AC power supply - PSA2 input | 100 - 240 VAC, 50 - 60Hz |
| AC power supply - PSA2 output | 19VDC, 2.6 Amp |
| Mobile power adaptor - MPA1 input | 13.5 VDC nominal. 12 to 15.5 VDC , 7.0 Amp |
| Mobile power adaptor - MPA1 output | 19.2 DVC, 3.91 Amp |
| Transcend P8 Multi-night Battery | 14.4 VDC, 5,200 mAH |
| Transcend P4 Overnight Battery | 14.4 VDC, 2,600 mAH |
| Working pressure range | 4 to 20 cm H2O |
| Pressure limit | 30 cm H2O |
| Sound power level (@ 10cm H2O pressure, static) | 34.6 dB |
| Sound pressure level (@ 10cm H2O pressure, static) | 26.6 dB |
| Resistance to flow of the mask and application accessories @ 50 l/min | 0.140 hPa (cm H2O) |
The working pressure range is 4 to 20 cm H2O.
The dimensions are 6.1 in x 3.5 in x 2.8 in (15.4 cm x 8.9 cm x 7.0 cm).
The weight is less than 1.0 lbs (454 g).
The device is contraindicated in patients with bullous lung disease, pathologically low blood pressure, pneumothorax or pneumomediastinum, and pneumocephalus has been reported in some users using nasal PAP.