Rotronic RMS-TCD-S Digital Temperature Probe Short Instruction Manual

This document is a short instruction manual for the Rotronic RMS-TCD-S digital temperature probe, which introduces the connection, installation, adjustment and calibration steps.


Rotronic RMS-TCD-S Digital Temperature Probe Short Instruction Manual - cover page
Brand
Rotronic
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
885 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
6CD8B8326FBFA3BE841CF9BC1BA415C3

About this manual

The document provides brief instructions for connecting the Rotronic RMS-TCD-S digital temperature probe to a transmitter or cable. It emphasizes aligning catches correctly and tightening the knurled nut by hand to ensure the IP protection rating is maintained.

For mounting, the manual states that the probe should be completely immersed in the environment being measured. It also notes that probes are adjusted at the factory but recommends recalibration after one year of use.

What's inside

  • General Description
  • Connection
  • Mounting
  • Adjustment / Calibration
  • Technical Data

Specifications

Measurement range-190...200°C
Accuracy±0,25 °C bei -50…85 °C
Long term stability±0.5 °C/year
T6390 s
Housing materialPolycarbonate
Housing diameterØ15 mm
Housing length110 mm
IP Rating (Housing)IP65
Sensor materialStainless steel - austenitic - 1.4571 (316Ti)
Sensor diameterØ3 mm
Sensor length50 mm
IP Rating (Sensor)IP68

Frequently asked questions

When should the probe be recalibrated?

We recommend a calibration after 1 year of use.

How do you connect the RMS-TCD-S probe?

Plug it onto the mating connector and tighten the knurled nut by hand.