Fluke 1550B MegOhmMeter Calibration Manual

1550B MegOhmMeter Calibration Manual is Fluke company's product warranty manual, which introduces the warranty scope, warranty period, warranty service and other content of the product


Fluke 1550B MegOhmMeter Calibration Manual - cover page
Brand
Fluke
Category
Industrial Supply
Document type
Service Manual
Language
English
Pages
28
File format
PDF
File size
606 KB
Published
15 October, 2023
Updated
15 October, 2023
Model
1550B
MD5 checksum
C883B767C0D79ED94060DB1927A3697B

About this manual

This document serves as the official calibration manual for the Fluke 1550B MegOhmMeter. It provides detailed instructions for technicians to verify instrument performance, perform necessary adjustments, and conduct battery replacement or disassembly procedures. The manual covers specific tests for insulation accuracy, output voltage, and short circuit current, as well as interface connection and instrument setup.

What's inside

  • Introduction
  • Safety Information
  • Required Equipment
  • Performance Test Procedure
  • Adjustment Procedure
  • Instrument Setup
  • Battery Replacement Procedure
  • Disassembly
  • Specifications
  • General Specification

Specifications

Display Size75 mm x 105 mm
Power Source12 V lead-acid rechargeable battery
Charger Input85 V to 250 V ac 50/60 Hz 20 VA
Dimensions170 mm x 242 mm x 330 mm
Weight3.6 kg (7.94 lbs.)
Operating Temperature-20 °C to 50 °C
Storage Temperature-20 °C to 65 °C
Altitude2000 m
Enclosure SealingIP40
Input Overload Protection600 V AC
Safety ComplianceEN 61010, EN 61557 Parts 1 and 2

Frequently asked questions

What is the warranty period for Fluke products?

The warranty period is two years and begins on the date of shipment.

What is the accuracy of the insulation resistance measurement for the 1550B?

The accuracy is +/- 5% for ranges from 200 kQ to 5 GQ, and +/- 20% for ranges from 5 GQ to 50 GQ.

What is the typical battery charge capability?

The typical battery charge capability is 5 seconds per uF for capacitive loads.

What is the operating temperature range for the 1550B?

The operating temperature range is -20 °C to 50 °C (-4 °F to 122 °F).