Shimadzu UV-2600 UV Spectrophotometer Operation Manual

The UV-2600 UV spectrophotometer is manufactured by Shimadzu in Japan and is combined with the powerful operating software UVProbe for easy and convenient operation. The main technical parameters include wavelength range, resolution, wavelength accuracy, spectral bandwidth range, etc.


Shimadzu UV-2600 UV Spectrophotometer Operation Manual - cover page
Brand
Shimadzu
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
15
File format
PDF
File size
1.1 MB
Published
30 November, 2022
Updated
27 September, 2023
Model
UV-2600
MD5 checksum
0288F85D888406E10BD4B30C9FA1A7BC

About this manual

The Shimadzu UV-2600 UV Spectrophotometer is a laboratory instrument manufactured by Shimadzu, designed for simple and convenient operation when combined with the UVProbe software. It features a double-beam measurement method and a Chopper-Turner monochromator, suitable for various analytical tasks in laboratory settings.

The document details the instrument's structure, operating procedures including startup and software connection, and various measurement modes such as photometric, spectral, and kinetic analysis. It also covers data saving, export methods, and common fault troubleshooting. This manual is written in Chinese.

What's inside

  • Instrument Overview
  • Technical Specifications
  • Operating Conditions
  • Startup Procedure
  • Software Connection
  • Photometric Measurement
  • Spectral Measurement
  • Kinetic Analysis
  • Data Saving and Export
  • Troubleshooting

Specifications

Wavelength Range185nm~900nm
Resolution0.1nm
Wavelength Accuracy±0.3nm
Slit Widths0.1, 0.2, 0.5, 1, 2, 5nm
Measurement MethodDouble-beam
Stray LightBelow 0.005%
DetectorPhotomultiplier R-928
Measuring Range-5~5Abs
Cuvette Path Length1cm
Light Source50W Halogen Lamp, Deuterium Lamp
MonochromatorChopper-Turner Monochromator
Main VoltageAC100V~240V
Main Frequency50~60Hz

Frequently asked questions

What are the operating temperature and humidity conditions for the UV-2600?

The operating temperature should be between 15 and 35 degrees Celsius, and the humidity should be between 30% and 80%.

How long does the instrument self-check take after startup?

The self-check takes approximately 5 minutes, indicated by a series of beeps and a change in indicator light status.

What is the recommended procedure for baseline correction?

It is recommended to perform baseline correction after the instrument has been preheated for 1 hour to ensure stability.