The UV-2600 UV-Vis spectrophotometer is manufactured by Shimadzu in Japan and is combined with the powerful UVProbe software for easy operation. This spectrophotometer features a wide wavelength range, high resolution, dual-beam measurement, and excellent detector performance.
This manual provides operating instructions for the Shimadzu UV-2600 UV-Vis Spectrophotometer. It details the instrument's specifications, including a wavelength range of 185nm to 900nm and a resolution of 0.1nm. The guide explains how to connect the hardware with the UVProbe software for data acquisition. This manual is written in Chinese.
| Wavelength Range | 185nm~900nm |
|---|---|
| Resolution | 0.1nm |
| Wavelength Accuracy | ±0.3nm |
| Slit Widths | 0.1, 0.2, 0.5, 1, 2, 5nm |
| Detector Type | Photomultiplier Tube R-928 |
| Light Source | 50W Halogen Lamp, Deuterium Lamp |
| Operating Voltage | AC100V~240V |
The instrument should be operated at a temperature between 15°C and 35°C with relative humidity between 30% and 80%.
The self-check takes approximately 5 minutes, indicated by a series of beeps and indicator light changes.
If detection shows an abnormality, close the power supply (off), ensure no items block the light path in the sample chamber, and turn the power on again.