BD Biosciences Th1/Th2 Subgroup Monitoring Test Kit Operation Guide (Single Laser Calibur)

This document is an operation guide for the Th1/Th2 Subgroup Monitoring Test Kit on a single laser Calibur flow cytometer. The guide provides detailed instructions on setting up the machine and acquiring samples.


BD Biosciences Th1/Th2 Subgroup Monitoring Test Kit Operation Guide (Single Laser Calibur) - cover page
Brand
BD Biosciences
Document type
User Manual
Language
Chinese
Pages
3
File format
PDF
File size
524 KB
Published
26 September, 2022
Updated
28 September, 2023
MD5 checksum
51C4A60DF94E9D02FA8B132FCF5BFA20

About this manual

This document serves as an operation guide for using the Th1/Th2 Subgroup Monitoring Test Kit in conjunction with a Single Laser Calibur flow cytometer. It details the initial machine condition settings required before analysis begins.

The procedure includes specific steps for preparing calibration beads and buffers, adjusting voltage thresholds (FSC, SSC) and fluorescence channels (FL1, FL2, FL3), and setting compensation parameters to ensure accurate data collection. This manual is written in Chinese.

What's inside

  • Machine Condition Setup
  • Calibration Bead Preparation
  • Voltage Adjustment
  • Compensation Settings
  • Sample Acquisition

Frequently asked questions

What voltage setting is required for the FSC threshold?

The FSC threshold must be adjusted to a value of 650.

How many events should be collected for each bead population?

The system should be set to collect approximately 300 beads for each population.

What is the required resolution value for data acquisition?

The Resolution value must be adjusted to 1024.