About this manual

The HI 96831 is a rugged, portable, water-resistant digital refractometer designed to determine the % volume and freezing point of ethylene glycol-based coolants or antifreeze. It utilizes Automatic Temperature Compensation (ATC) and features an automatic shut-off function after three minutes of non-use.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Principle of Operation
  • Measurement Units
  • Functional Description
  • Calibration Procedure
  • Measurement Procedure
  • Error Messages

Specifications

Range (% Volume)0 to 100 %
Range (Freezing Point °C)-50 to 80 °C
Resolution (% Volume)0.1%
Accuracy (% Volume)+/-0.2 %
Minimum Sample Volume100 µl
Battery Life5000 readings
Enclosure RatingIP 65

Frequently asked questions

How often should the instrument be calibrated?

Calibration should be performed daily, before measurements are made, when the battery has been replaced, between a long series of measurements, or if environmental changes have occurred.

What solution is used for zeroing the instrument?

Distilled or deionized water must be used to fill the sample well for calibration.

What happens if the temperature exceeds 40°C during measurement?

The 'ATC' tag will blink and automatic temperature compensation is disabled.