Hanna Instruments HI 96717 Phosphate High Range ISM Instruction Manual

This document describes an ion selective meter called HI 96717, which features GLP mode, calibration mode, zeroing function, etc.


Hanna Instruments HI 96717 Phosphate High Range ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
611 KB
Published
10 July, 2012
Updated
29 September, 2023
MD5 checksum
72A24A460D555414F609945095D4F81B

About this manual

This document is the instruction manual for the Hanna Instruments HI 96717 Phosphate High Range ISM (In-Situ Meter). It provides preliminary examination steps, lists included accessories such as sample cuvettes and a 9V battery, and details technical specifications including range, resolution, accuracy, and environmental conditions. The meter utilizes an amino acid method adapted from standard water examination methods.

What's inside

  • Preliminary examination
  • Technical specifications
  • Functional description
  • Errors and warnings
  • Measurement procedure
  • Interferences
  • Validation and Calibration procedures
  • Battery management

Specifications

Range10 to 30.0 mg/l
Resolution0.1 mg/l
Accuracy+/-1 mg/L +/-4% of reading @ 25°C
Typical RSD Dev.0.2 mg/L
Light SourceTungsten lamp
Light DetectorSilicon Photocell with narrow band interference filter @ 525 nm
MethodAmino Acid Method, adapted from Standard Method for the Examination of Water and Wastewater.
Operating Temperature0 to 50°C (32 to 122°F)
Humidity<95% RH non-condensing
Battery Type1x9 volt
Auto-Shut off (Measurement)After 10' of non-use
Dimensions192 x 102 x 67 mm (7.6x4x 2.6")
Weight290 g (10 oz)

Frequently asked questions

What is the measurement range of the HI 96717?

The instrument measures phosphate concentrations in a range from 10 to 30.0 mg/l.

How long does the battery last?

One fresh battery lasts for around 750 measurements, depending on the light level.

What happens if the sample concentration is too high?

A flashing value of the maximum concentration indicates an over range condition; you must dilute the sample and re-run the test.

Can I use standard solutions other than Hanna CAL CHECK Standards for calibration?

No, you should not validate or calibrate with other standards as it will result in erroneous readings.

Download PDF (611 KB)

More Hanna Instruments Manuals


View all Hanna Instruments manuals →