Hanna Instruments HI 93717 Phosphate High Range ISM Instruction Manual

This manual provides instructions for the use of the HI 93717 Phosphate High Range ISM. It covers the instrument's features, operation, and maintenance.


Hanna Instruments HI 93717 Phosphate High Range ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
474 KB
Published
10 July, 2012
Updated
01 October, 2023
MD5 checksum
B5203D34553B53A8C91F87630614C6C3

About this manual

The HI 93717 is a portable instrument designed to measure phosphate (PO4) content in water and wastewater within a range of 0.0 to 30.0 mg/L. It utilizes an exclusive positive-locking system for cuvet placement and features an auto-shut off function after 10 minutes of non-use. The device operates using a light source of 470 nm and a silicon photocell to detect the blue tint caused by the reaction between phosphate and reagents.

The manual details the step-by-step measurement procedure, including zeroing the instrument, adding specific reagents (HI 93717A-0 and HI 93717B-0), and reading the results. It also lists potential interferences such as sulfide, high chloride, calcium, magnesium, and ferrous iron, along with tips for ensuring accurate measurements and instructions for battery replacement.

What's inside

  • Warranty
  • General Description
  • Specifications
  • Display Code Guide
  • Measurement Procedure
  • Interferences
  • Tips for Accurate Measurement
  • Battery Replacement
  • Accessories
  • Safety Recommendations

Specifications

Range0.0 to 30.0 mg/L
Resolution0.1 mg/L
Accuracy+1 mg/L +4% of reading
Light SourceLight Emitting Diode @ 470 nm
Operating Temperature0 to 50°C (32 to 122°F)
Battery Type1 x 9 volt
Auto-Shut offAfter 10 minutes of non-use
Dimensions80 x 83 x 46 mm
Weight290 g (10 oz)

Frequently asked questions

What is the measurement range for the HI 93717?

The meter measures phosphate content in the 0.0 to 30.0 mg/L (ppm) range.

What reagents are required for testing?

The instrument requires HI 93717A-0 Phosphate HR Reagent A and HI 93717B-0 Phosphate HR Reagent B.

What causes interference during measurement?

Interference may be caused by sulfide, chloride above 150000 mg/L, calcium above 10000 mg/L, magnesium above 40000 mg/L, ferrous iron above 100 mg/L, or large amounts of color or suspended matter.

How long does the battery last?

The battery life is approximately 40 hours.

Download PDF (474 KB)