Hanna Instruments HI 84184 ISE Fluoride Meter Instruction Manual

This document is an instruction manual for the HI 84184 ISE Fluoride Meter, providing information on the correct use of the instrument, technical specifications, functional description, startup process, measurement procedures, etc.


Hanna Instruments HI 84184 ISE Fluoride Meter Instruction Manual - cover page
Brand
Hanna Instruments
Document type
User Manual
Language
English
Pages
20
File format
PDF
File size
625 KB
Published
10 July, 2012
Updated
28 September, 2023
MD5 checksum
2D6EFAC9B21A31D895CA80D4C28E5C95

About this manual

The HI 84184 is a low-cost, easy-to-use fluoride ion selective electrode (ISE) instrument designed to perform automatic analysis with built-in calculations. It utilizes the double standard addition method to determine fluoride concentration in samples such as wine, displaying results immediately in mg/L (ppm). The device features an automatic temperature compensation system and includes a powerful algorithm to analyze electrode response shapes.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Principle of Operation
  • Functional Description
  • Start Up
  • Guide to Display Codes
  • Measurement Procedure
  • Electrode Conditioning and Maintenance

Specifications

Range0.0 to 5.0 mg/L (ppm) F
Resolution0.1 mg/L (ppm) F
Accuracy+5% of reading
Stirring speed500 rpm
Power supply220V/50Hz; 10VA

Frequently asked questions

What method does the HI 84184 use for analysis?

The instrument uses the double standard addition method to determine unknown concentration and electrode slope factor simultaneously.

What is the acceptable range for the computed slope?

The acceptable slope range is between 30% and 140%, with a default of 56 mV/decade.