Hanna Instruments HI 84430 Total Acidity Minititrator & pH Meter Instruction Manual

This document is an instruction manual for the HI 84430 TOTAL ACIDITY MINITITRATOR & pH METER. It provides information about the specifications, usage, calibration procedures, and troubleshooting guide of the instrument.


Hanna Instruments HI 84430 Total Acidity Minititrator & pH Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
25
File format
PDF
File size
447 KB
Published
10 July, 2012
Updated
29 September, 2023
MD5 checksum
800BACF9A3CA4CF57166FF222305763B

About this manual

The HI 84430 is an easy-to-use instrument designed for the rapid and accurate analysis of Total Titratable Acidity in water. It functions as both a titrator and a pH meter, utilizing potentiometric end-point determination to eliminate subjective factors associated with color indicators.

The device features a clear user interface, a dedicated HELP key, and automatic pump operation initiated by a single START key press. It includes built-in algorithms for analyzing pH responses and calculating acidity results in units such as mg/L or meq/L as CaCO3.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Principle of Operation
  • Functional and Physical Description
  • Titrator Startup
  • Setup Configuration Menu
  • Electrode Preparation
  • Titration Procedure
  • Troubleshooting Guide

Specifications

pH Range-2.0 to 16.0 pH / -2.00 to 16.00 pH
pH Resolution0.1 pH / 0.01 pH
Titratable Acidity Range (Low)15.0 - 500.0 mg/L as CaCO3
Titratable Acidity Range (High)400 - 4000 mg/L as CaCO3
Pump Debit0.5 ml/min
Stirring Speed600 rpm
Temperature Range-20.0 to 120.0 °C (-4.0 to 248.0 °F)
Temperature Accuracy+/- 0.4 °C without probe error

Frequently asked questions

What is the pH end point for total acidity determination?

Total acidity determination uses an alkaline reagent titrated to a pH 8.3 endpoint.

What is the pH end point for strong acidity determination?

Strong acidity is determined by titrating water samples with an alkaline reagent to a pH 3.7 endpoint.