This is a device for measuring the peroxide value of edible oils. It is easy to operate and has a measurement range of 0.0 to 25.0 meq O2/kg, a resolution of 0.5 meq O2/kg, and an accuracy of ±0.5 meq O2/kg.
The HI 83730 is an auto-diagnostic portable microprocessor meter designed for the fast and simple analysis of peroxides in oil. It utilizes an advanced optical system with a tungsten lamp and a narrow band interference filter at 466 nm, complying with the EC 2568/91 method. The instrument features an auto-diagnostic function that ensures optimal measurement conditions by automatically adjusting light levels and controlling lamp temperature.
| Range | 0.0 to 25.0 meq O/kg |
|---|---|
| Resolution | 0.5 meq O/kg |
| Accuracy | + 0.5 meq O/kg |
| Light source | tungsten lamp with narrow band interference filter @ 466 nm |
| Light detector | silicon photocell |
| Environment | 0 to 50 °C (82 to 122°F), max 95% RH |
| Power supply | 4x1.5 ¥ AA or adapter 12 Vdc |
| Auto shut-off | after 15 min of non use |
| Weight/dimensions | 512 g; 224 x 87 x 77 mm |
The measurement range is 0.0 to 25.0 meq O/kg.
The instrument uses the adaptation of the EC 2568/91 method and following amendments.
The instrument automatically shuts off after 15 minutes of non-use.
A blinking 'tex.' will appear on the display to indicate that the batteries need to be replaced.