This document describes the Agilent Operational Qualification/Performance Verification (OQ/PV) procedures for the 1100 Series Liquid Chromatograph, which use internal tests and analytical methods (certified samples) to evaluate instrument performance and verify that the instrument is operating within acceptable specifications for both Agilent and you.
This manual details the procedures for Operational Qualification (OQ) and Performance Verification (PV) of the Agilent 1100 Series Liquid Chromatograph. It utilizes internal tests and analytical methods with certified samples to evaluate instrumental performance and verify that the instrument functions within acceptable specifications for both Agilent and the user. The ChemStation software (Revision 4.06.03 or later) is used to automate these semi-automated and automated qualification procedures for the HPLC modules. This manual is written in Spanish.
| Qualification Time | Normally one day |
|---|---|
| Applicable Instruments | New instruments and those maintained/repairs under Agilent specifications |
The ChemStation software is adequate for automated and semi-automated OQ/PV procedures for the HPLC Series 1100 modules.
The processes can usually be completed in one day to minimize downtime.
Permitted hardware includes LC 1100 pumps, injection systems, and various detectors.