Agilent 5973/5975 MSD EI/CI Source Cleaning Installation Manual

This overview describes a simple approach to cleaning and re-installing the 5973 and 5975 ion electron impact (EI) inert and standard sources and the chemical ionization (CI) source.


Agilent 5973/5975 MSD EI/CI Source Cleaning Installation Manual - cover page
Brand
Agilent
Document type
Service Manual
Language
English
Pages
6
File format
PDF
File size
805 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
FCBC2D6C902EB19A204C3BF3E5F711F1

About this manual

This manual provides a specific approach to cleaning and reinstalling the EI and CI ion sources for Agilent 5973 and 5975 MSD instruments. It details a multi-step process involving mechanical polishing with alumina MICROGRIT, followed by sonication in deionized water, methanol, acetone, and hexane to remove contaminants without damaging source surface chemistry.

What's inside

  • Introduction
  • Precautions
  • Cleaning Procedure
  • Reinstallation Procedure
  • Analyzer Baking
  • Automated Baking and Performance Checkout

Frequently asked questions

What solvents are recommended for the cleaning process?

The procedure recommends using deionized water, methanol (pesticide or HPLC grade), acetone (pesticide or HPLC grade), and hexane (pesticide or HPLC grade).

What is the recommended baking temperature for the ion source?

The manual recommends setting the ion source bake temperature to 300 °C.

How long should the parts be sonicated in each solvent?

Parts should be sonicated for about 5 minutes in deionized water, methanol, acetone, and hexane.