Agilent GS-OxyPLOT PLOT Column Product Information

Introduced the characteristics and uses of GS-OxyPLOT column


Agilent GS-OxyPLOT PLOT Column Product Information - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
49 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
22365BBA57923A969C33BD0AC3A2E69F

About this manual

The Agilent GS-OxyPLOT is a porous layer open tubular (PLOT) column featuring a proprietary, salt-deactivated adsorbent stationary phase. It offers high chromatographic selectivity specifically for low molecular weight oxygenated hydrocarbons and serves as an appropriate replacement for Varian's CP-LowOx column.

The document provides technical specifications including temperature limits (0 °C minimum, 350 °C maximum oven program), conditioning instructions, and sample loading limitations. It lists specific ASTM methods the column is designed for and provides a table of suitable oxygenated compounds such as methanol, ethanol, and various ethers and aldehydes.

What's inside

  • Introduction
  • Analysis of Oxygenated Hydrocarbons
  • Technical Overview
  • ASTM Standardized Methods
  • Examples of Suitable Compounds
  • Ordering Information

Specifications

Stationary Phase Typeproprietary, salt deactivated adsorbent
Minimum Temperature Limit0 °C
Isothermal Maximum Temperature Limit300 °C
Oven Program Maximum Temperature350 °C
Conditioning Temperature300 °C for at least 3 hours

Frequently asked questions

What is the recommended conditioning procedure for a new GS-OxyPLOT column?

When first installed, the column should be conditioned at 300 °C for at least 3 hours.

What is the recommended idle temperature setting for the GC oven?

It is recommended to set the GC oven to an isothermal temperature of 220 °C with normal carrier gas flow when the column is installed but idle.

What are the maximum carbon counts for unsaturated and aromatic hydrocarbons?

Injection should be limited to organic compounds with 11 carbons or less to prevent the column from fouling.