Agilent Eclipse XDB-C18 RRHT Threaded Column Datasheet

This document is the datasheet of Agilent Eclipse XDB-C18 RRHT Threaded Column


Agilent Eclipse XDB-C18 RRHT Threaded Column Datasheet - cover page
Brand
Agilent
Category
Other
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
83 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
942490795DA8962BFD1D29615E23B4FB

About this manual

This datasheet describes the Agilent Eclipse XDB-C18 RRHT threaded columns, which are specially designed for higher pressure operation up to 600 bar. The columns utilize eXtra-Dense Bonding (XDB) technology and double endcapping to protect the ultra-pure silica support from dissolution in mobile phases of intermediate and higher pH. They are packed with a high-performance microparticulate C18 packing for high-speed reversed-phase HPLC separation of acidic, basic, and highly polar compounds.

What's inside

  • General Description
  • Operating Conditions
  • Column Characteristics
  • Safety Considerations
  • Operational Guidelines
  • Mobile Phase Selection
  • Applications
  • Column Care
  • Storage Recommendations

Specifications

Maximum operating pressure600 bar (9000 psi)
Maximum operating temperature60 °C
Optimum pH rangepH 3-8
Particle size1.8 um
Controlled pore size80A

Frequently asked questions

What is the maximum operating pressure for these columns?

The maximum operating pressure is 600 bar (9000 psi).

What is the recommended pH range for optimum performance?

Optimum lifetime and performance are obtained at pH 3-8.

Can these columns be used with basic compounds?

Yes, they are especially suited for separating basic compounds.