This document introduces the photometric accuracy of the Cary 4000, 5000 and 6000i spectrophotometers with the Agilent DRA.
This document describes a method to determine the photometric accuracy of the whole system when using diffuse reflectance accessories (DRAs) with in-built detectors on Agilent New Generation Cary 4000, 5000, and 6000i spectrophotometers. The test utilizes a certified potassium dichromate solution as a standard.
The measured values must be within an acceptable error of 0.01 absorbance.
The scan rate must be set to 60.0 nm/min with a data interval of 1.0 nm.