Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet

This document introduces the photometric accuracy of the Cary 4000, 5000 and 6000i spectrophotometers with the Agilent DRA.


Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet - cover page
Brand
Photometric
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
159 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
44130BD7A314252110EF14D4FC426492

About this manual

This document describes a method to determine the photometric accuracy of the whole system when using diffuse reflectance accessories (DRAs) with in-built detectors on Agilent New Generation Cary 4000, 5000, and 6000i spectrophotometers. The test utilizes a certified potassium dichromate solution as a standard.

What's inside

  • Introduction
  • Materials
  • Method
  • Results
  • Conclusion

Frequently asked questions

What is the acceptable error limit for the photometric accuracy test?

The measured values must be within an acceptable error of 0.01 absorbance.

What is the required scan rate for the measurement procedure?

The scan rate must be set to 60.0 nm/min with a data interval of 1.0 nm.