Photometric 6000i Manuals

ManualLib has 2 documents for the Photometric 6000i: 1 product leaflet, 1 datasheet. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 1Datasheet 1

Key facts

Model
6000i
Documents
2
Languages
EN
Last added
2012-07-04

Start here

Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet

Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet

Product Leaflet 4 pages · 159 KB · EN

This document describes a method to determine the photometric accuracy of the whole system when using diffuse reflectance accessories (DRAs) with in-built detectors on Agilent New Generation Cary 4000, 5000, and 6000i spectrophotometers. The test utilizes a certified potassium dichromate solution as a standard.

Read online

All Photometric 6000i documents by type

Datasheet 1

Ratings, dimensions and ordering codes, for selection and design work.

Questions answered in these manuals

What is the acceptable error limit for the photometric accuracy test?

The measured values must be within an acceptable error of 0.01 absorbance.

From: Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet

What is the required scan rate for the measurement procedure?

The scan rate must be set to 60.0 nm/min with a data interval of 1.0 nm.

From: Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet

What part number is required for routine linearity tests?

Routine linearity tests should be performed using the Linear Neutral Density Kit, part number 9910056100.

From: Agilent Cary 4000/5000/6000i Spectrophotometer Photometric Linearity Data Sheet

What is the maximum absorbance level demonstrated in the standard method?

Using the standard method parameters, photometric linearity of up to 9 absorbance can be demonstrated.

From: Agilent Cary 4000/5000/6000i Spectrophotometer Photometric Linearity Data Sheet