Start here

Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet
This document describes a method to determine the photometric accuracy of the whole system when using diffuse reflectance accessories (DRAs) with in-built detectors on Agilent New Generation Cary 4000, 5000, and 6000i spectrophotometers. The test utilizes a certified potassium dichromate solution as a standard.
Read onlineAll Photometric Cary 4000 documents by type
Ratings, dimensions and ordering codes, for selection and design work.
Questions answered in these manuals
What is the acceptable error limit for the photometric accuracy test?
The measured values must be within an acceptable error of 0.01 absorbance.
From: Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet
What is the required scan rate for the measurement procedure?
The scan rate must be set to 60.0 nm/min with a data interval of 1.0 nm.
From: Agilent Cary 4000/5000/6000i DRA Photometric Accuracy Data Sheet
What part number is required for routine linearity tests?
Routine linearity tests should be performed using the Linear Neutral Density Kit, part number 9910056100.
From: Agilent Cary 4000/5000/6000i Spectrophotometer Photometric Linearity Data Sheet
What is the maximum absorbance level demonstrated in the standard method?
Using the standard method parameters, photometric linearity of up to 9 absorbance can be demonstrated.
From: Agilent Cary 4000/5000/6000i Spectrophotometer Photometric Linearity Data Sheet