Agilent 6890 Analysis of Trace Sulfur Compounds in Beverage Grade Carbon Dioxide Application Note

Describes an inert valving system for the analysis of ppb level volatile sulfur compounds in food grade carbon dioxide using the Agilent 6890.


Agilent 6890 Analysis of Trace Sulfur Compounds in Beverage Grade Carbon Dioxide Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
154 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
5ED871D3DD7F946B912BBD0C85864777

About this manual

The document details a method for analyzing ppb-level volatile sulfur compounds in food grade carbon dioxide using an inert valving system on the Agilent 6890 gas chromatograph. It highlights the importance of strict adherence to sample introduction hardware recommendations, specifically using Silcosteel treated components, to prevent contamination and ensure reliable operation for low-level detection.

The note presents experimental results using three sulfur-selective detectors: the Flame Photometric Detector (FPD), Atomic Emission Detector (AED), and Sulfur Chemiluminescent Detector (SCD). It compares their sensitivity and suitability for detecting specific impurities like hydrogen sulfide, carbonyl sulfide, and dimethyl sulfide at concentrations down to 25 ppb.

What's inside

  • Introduction
  • Experimental
  • Detector Settings
  • Results
  • Conclusions

Specifications

Column105 m x 0.53 mm x 5 um DB-1
Gas sampling valve loop1 cc
VI Inlet Temperature150 °C
FPD Hydrogen Flow50 mL/min
AED Cavity Temperature245 °C
SCD Furnace Temperature803 °C

Frequently asked questions

Which detectors are used for sulfur analysis in this application?

The study uses the Agilent Flame Photometric Detector (FPD), Atomic Emission Detector (AED), and Sievers Sulfur Chemiluminescent Detector (SCD).

What material is recommended for the sample introduction system?

All lines and the sample loop should be Silcosteel treated to provide the necessary inertness for low level sulfur analysis.