Agilent ScreenPlex System Quick Guide introduces the assay principles, composition and use of the Agilent 2200 TapeStation system. The system can be used for simpler, faster and more reliable sample electrophoresis.
This document serves as a quick guide for the Agilent 2200 TapeStation ScreenPlex System (p/n G2966AA). It outlines the system components, including the ScreenTape DS12 reagents and compatible assay kits from Seegene Inc., designed for electrophoretic separation of double-stranded DNA species between 100 bp to 1000 bp. The guide explains how the system loads, separates, images, analyzes, and presents results after sample amplification.
The system detects fluorescently stained double stranded DNA products between 100 base pairs (bp) to 1000 bp.
Reagent vials and ScreenTape must be stored at 2—8 °C. Used tape can be stored upright at this temperature for a maximum of 2 weeks if fewer than 16 lanes were run.
The TapeStation software will not launch without the TestKitSuite.xml file, which must be obtained from the PCR Test Kit manufacturer.