Hewlett-Packard 8648A/B/C/D Error 605: ALC out-of-lockdivhetmainband

Hewlett-Packard 8648A/B/C/D Signal Generator · taken from the manufacturer’s manual, pages cited

Error code605

ALC out-of-lockdivhetmainband means either the A6 output has failed, or the signal from A5 synthesizer has failed or the A4 reference has failed. If only the het band has failed it is possible that the A4 reference has failed. The manual's first action is: “Check the input to the A6 output from the A5 synthesizer and the A4 reference.”

Model
Hewlett-Packard 8648A/B/C/D
Display shows
ALC out-of-lockdivhetmainband (605)
Type
Error code
Source
Manual p. 228

What 605 means

ALC out-of-lockdivhetmainband p. 228

Common causes

  • Either the A6 output has failed, or the signal from A5 synthesizer has failed or the A4 reference has failed. If only the het band has failed it is possible that the A4 reference has failed. p. 228

What to do

  1. Check the input to the A6 output from the A5 synthesizer and the A4 reference. p. 228

Before you service the product, read the safety instructions in the manual p. 277.

If 605 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.

This page summarises what the manufacturer’s manual says about 605; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.