Outputprelevelcalrestorefailure means there is either a failure in writing data to RAM or in reading data from the calibration ROM. The manual's first action is: “The calibration ROM can be verified by replacing the A6 assembly. The calibration ROM is contained in the assembly. If the failure is still present after replacing the A6 assembly, replace the A3 assembly, which contains the RAM.”
- Model
- Hewlett-Packard 8648A/B/C/D
- Display shows
- Outputprelevelcalrestorefailure (517)
- Type
- Error code
- Source
- Manual p. 227
What 517 means
Outputprelevelcalrestorefailure p. 227
Common causes
- There is either a failure in writing data to RAM or in reading data from the calibration ROM. p. 227
What to do
- The calibration ROM can be verified by replacing the A6 assembly. The calibration ROM is contained in the assembly. If the failure is still present after replacing the A6 assembly, replace the A3 assembly, which contains the RAM. p. 227
Before you service the product, read the safety instructions in the manual p. 277.
If 517 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.
This page summarises what the manufacturer’s manual says about 517; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.