Hewlett-Packard 8648A/B/C/D Error 614: 400 Hz modulation source failed

Hewlett-Packard 8648A/B/C/D Signal Generator · taken from the manufacturer’s manual, pages cited

Error code614

The 400 Hz modulation source failed code indicates either the 200 kHz reference signal from the A4 module has failed or the A3 has failed. The manual's first action is: “Check the 200 kHz reference output of the A4 module at the input to the A3 board.”

Model
Hewlett-Packard 8648A/B/C/D
Display shows
400 Hz modulation source failed (614)
Type
Error code
Source
Manual p. 229

What 614 means

400 Hz modulation source failed p. 229

Common causes

  • Either the 200 kHz reference signal from the A4 module has failed, or the A3 has failed. p. 229

What to do

  1. Check the 200 kHz reference output of the A4 module at the input to the A3 board. p. 229

Before you service the product, read the safety instructions in the manual p. 277.

If 614 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.

This page summarises what the manufacturer’s manual says about 614; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.