The 400 Hz modulation source failed code indicates either the 200 kHz reference signal from the A4 module has failed or the A3 has failed. The manual's first action is: “Check the 200 kHz reference output of the A4 module at the input to the A3 board.”
- Model
- Hewlett-Packard 8648A/B/C/D
- Display shows
- 400 Hz modulation source failed (614)
- Type
- Error code
- Source
- Manual p. 229
What 614 means
400 Hz modulation source failed p. 229
Common causes
- Either the 200 kHz reference signal from the A4 module has failed, or the A3 has failed. p. 229
What to do
- Check the 200 kHz reference output of the A4 module at the input to the A3 board. p. 229
Before you service the product, read the safety instructions in the manual p. 277.
If 614 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.
This page summarises what the manufacturer’s manual says about 614; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.