Hewlett-Packard 8648A/B/C/D Error 611: 200kHzreference missing at synth

Hewlett-Packard 8648A/B/C/D Signal Generator · taken from the manufacturer’s manual, pages cited

Error code611

The 200kHzreference missing at synth code indicates either the A4 module has failed to output the 200 kHz reference signal or the A5 module is failing to detect the signal. The manual's first action is: “Check the 200 kHz reference output of the A4 module at the input to the A5 module.”

Model
Hewlett-Packard 8648A/B/C/D
Display shows
200kHzreference missing at synth (611)
Type
Error code
Source
Manual p. 228

What 611 means

200kHzreference missing at synth p. 228

Common causes

  • Either the A4 module has failed to output the 200 kHz reference signal, or the A5 module is failing to detect the signal. p. 228

What to do

  1. Check the 200 kHz reference output of the A4 module at the input to the A5 module. p. 228

Before you service the product, read the safety instructions in the manual p. 277.

If 611 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.

This page summarises what the manufacturer’s manual says about 611; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.