Hewlett-Packard 8648A/B/C/D Error 515: Freq Ext level cal restore failure

Hewlett-Packard 8648A/B/C/D Signal Generator · taken from the manufacturer’s manual, pages cited

Error code515

Freq Ext level cal restore failure means there is either a failure in writing data to RAM or in reading data from the calibration ROM. The manual's first action is: “The calibration ROM can be verified by replacing the A10 (freq ext) assembly. The calibration ROM is contained in the assembly. If the failure is still present after replacing the A10 assembly, replace the A3 assembly, which contains the RAM.”

Model
Hewlett-Packard 8648A/B/C/D
Display shows
Freq Ext level cal restore failure (515)
Type
Error code
Source
Manual p. 226

What 515 means

Freq Ext level cal restore failure p. 226

Common causes

  • There is either a failure in writing data to RAM or in reading data from the calibration ROM. p. 226

What to do

  1. The calibration ROM can be verified by replacing the A10 (freq ext) assembly. The calibration ROM is contained in the assembly. If the failure is still present after replacing the A10 assembly, replace the A3 assembly, which contains the RAM. p. 226

Before you service the product, read the safety instructions in the manual p. 277.

If 515 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.

This page summarises what the manufacturer’s manual says about 515; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.