Temperature cal restore failure means there is either a failure in writing data to RAM or in restoring data from the calibration ROM. The manual's first action is: “The calibration ROM can be verified by replacing the A7 assembly. The calibration ROM is contained in the module. If the failure is still present after replacing the A7 assembly, replace the A3 assembly, which contains the RAM.”
- Model
- Hewlett-Packard 8648A/B/C/D
- Display shows
- Temperature cal restore failure (510)
- Type
- Error code
- Source
- Manual p. 225
What 510 means
Temperature cal restore failure p. 225
Common causes
- There is either a failure in writing data to RAM or in restoring data from the calibration ROM. p. 225
What to do
- The calibration ROM can be verified by replacing the A7 assembly. The calibration ROM is contained in the module. If the failure is still present after replacing the A7 assembly, replace the A3 assembly, which contains the RAM. p. 225
Before you service the product, read the safety instructions in the manual p. 277.
If 510 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.
This page summarises what the manufacturer’s manual says about 510; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.