Fuji Electric FSC-1/FLD-1/FSD-1 Error E4: Range over

Ultrasonic flowmeter · taken from the manufacturer’s manual, pages cited

Error codeE4

Range over means the output signal is out of the adjustable range, or the analog input is abnormal. The manual's first action is: “-Check and re-set the measuring range. In case output signal is full message, set 'parameter setting of Analog Input/Output function to 'OFF'.”

Model
Fuji Electric FSC-1/FLD-1/FSD-1
Display shows
Range over (E4)
Type
Error code
Source
Manual p. 151

What E4 means

Range over p. 151

Common causes

  • Output signal is out of the adjustable range. p. 151
  • Analog input is abnormal. p. 151
  • In case of no recovery after above measures, contact the nearest Fuji sales office's distributor. In case Analog input signal is not needed, set the parameter setting of Analog Input/Output: function to 'OFF'. p. 151

What to do

  1. -Check and re-set the measuring range. In case output signal is full message, set 'parameter setting of Analog Input/Output function to 'OFF'. p. 151
  2. -Check the cable connection. -Check the input signal. -Perform re-calibration of analog input. -Turn the power off and on! p. 151

Before you service the product, read the safety instructions in the manual p. 6.

If E4 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Fuji Electric service.

This page summarises what the manufacturer’s manual says about E4; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Fuji Electric. Found a mistake? Report an error.