Fuji Electric FSC-1/FLD-1/FSD-1 Error E3: Measurement range error

Ultrasonic flowmeter · taken from the manufacturer’s manual, pages cited

Error codeE3

Measurement range error means the setting is out of the measurement range, or there is an error in the velocity distribution measurement calculation, or the particle reflectors or fluid volume is too small. The manual's first action is: “Set the action to be taken at the time of burn-out.”

Model
Fuji Electric FSC-1/FLD-1/FSD-1
Display shows
Measurement range error (E3)
Type
Error code
Source
Manual p. 119

What E3 means

Measurement range error p. 119

Common causes

  • Setting is out of a measurement range. Measurement Impossible. p. 119
  • Measurement Impossible due to Error in Measurement Calculation of velocity distribution. p. 119
  • Measurement Impossible due to too small a volume of particle reflectors or fluid. p. 119

What to do

  1. Set the action to be taken at the time of burn-out. p. 119
  2. -Check if the parameters or Site settings are OK. -Check if the sensor type is correctly selected. -Turn the power off and on! p. 119
  3. -Check if the parameters in Site setting are OK. -Check if the sensor type is correctly selected. -Turn the power off and on! p. 119
  4. -Check if the parameters or Site settings are OK. -Check if the sensor type is correctly selected. -Try to reduce the success rate value. p. 119

Before you service the product, read the safety instructions in the manual p. 6.

If E3 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Fuji Electric service.

This page summarises what the manufacturer’s manual says about E3; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Fuji Electric. Found a mistake? Report an error.