The detector's position in the room allows it to pass through a non-radioactive wall, which is not permitted. The manual's first action is: “Change the source-detector location parameters appropriately to fix this error. Then try again.”
- Model
- Canberra 9233653E V3.0
- Display shows
- DETECTOR IS OUT OF ROOM THROUGH THE SURFACE. (#1401)
- Type
- Error code
- Source
- Manual p. 436
What #1401 means
DETECTOR IS OUT OF ROOM THROUGH THE SURFACE. p. 436
Common causes
- In the 'room' template description, the source/detector location can be defined such that the detector penetrates through a non-radioactive wall of the room. But the detector must not penetrate through a radioactive wall. The above error message means that the detector is found to penetrate through a radioactive wall. p. 436
What to do
- Change the source-detector location parameters appropriately to fix this error. Then try again. p. 436
Before you service the product, read the safety instructions in the manual p. 526.
If #1401 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Canberra service.
This page summarises what the manufacturer’s manual says about #1401; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Canberra. Found a mistake? Report an error.
