E16 indicates the fan is not working, either due to an electrical failure or a physical obstruction. The manual's first action is: “Inspect the fan grid to check if the fan blade is rotating when the power supply is turned "ON". Make sure no foreign material has penetrated the fan vent. If problem persists, contact Bio-Rad.”
- Model
- Bio-Rad PowerPac HC
- Display shows
- E16
- Type
- Error code
- Source
- Manual p. 13
What E16 means
The fan is not working, either due to a electrical failure or physical obstruction. p. 13
Common causes
- The fan is not working, either due to a electrical failure or physical obstruction. p. 13
What to do
- Inspect the fan grid to check if the fan blade is rotating when the power supply is turned "ON". Make sure no foreign material has penetrated the fan vent. If problem persists, contact Bio-Rad. p. 13
If E16 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.
This page summarises what the manufacturer’s manual says about E16; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.