E02 indicates a short circuit detected, meaning over current greater than 3.0 A, accidental shorting of output leads, or shorting due to wrong connections. The manual's first action is: “• Insure all electrical connections are making good contact and the cables and wire electrodes are in good shape. • Verify buffer levels are appropriate.”
- Model
- Bio-Rad PowerPac HC
- Display shows
- E02
- Type
- Error code
- Source
- Manual p. 12
What E02 means
Short circuit detected (over current, load over current is greater than 3.0 A). • Accidental shorting of output leads. • Shorting due to wrong connections. p. 12
Common causes
- Short circuit detected (over current, load over current is greater than 3.0 A). • Accidental shorting of output leads. • Shorting due to wrong connections. p. 12
What to do
- • Insure all electrical connections are making good contact and the cables and wire electrodes are in good shape. • Verify buffer levels are appropriate. p. 12
If E02 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.
This page summarises what the manufacturer’s manual says about E02; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.