Bio-Rad PowerPac HC Error E09

Power supply · taken from the manufacturer’s manual, pages cited

Error codeE09

E09 indicates a change in load resistance detected, meaning the PowerPac HC detects drastic changes in resistance that may indicate failure of the cell's power leads or a loose output connection, loose output connections causing intermittent connection, or cells added or removed during the run. The manual's first action is: “Check and correct any potential resistance problem. Verify all electrical connections. Verify buffer levels are appropriate.”

Model
Bio-Rad PowerPac HC
Display shows
E09
Type
Error code
Source
Manual p. 13

What E09 means

Change in load resistance detected. The PowerPac HC detects drastic changes in resistance which may indicate failure of the cell's power leads or a loose output connection. p. 13

Common causes

  • Change in load resistance detected. The PowerPac HC detects drastic changes in resistance which may indicate failure of the cell's power leads or a loose output connection. p. 13
  • Loose output connections leading to intermittent connection to the loads. p. 13
  • Cells added or removed during the run. p. 13

What to do

  1. Check and correct any potential resistance problem. Verify all electrical connections. Verify buffer levels are appropriate. p. 13
  2. Certain applications exhibit intrinsic fluctuations in resistance that can trigger the Resistance Change error message. If this is the case, the Resistance Change detection feature can be deactivated to allow an uninterrupted completion of the run. See Section 3.4 for details on turning off RRCd. p. 13
  3. Caution: Deactivating this safety feature increases the possibility that a failure of the power leads or loose output connections will not be detected. p. 13

If E09 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.

This page summarises what the manufacturer’s manual says about E09; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.