Agilent E5061A/E5062A ENA Series Error 53: Log sweep requires 2 octave minimum span

ENA Series network analyzer · taken from the manufacturer’s manual, pages cited

Error code53

Log sweep requires 2 octave minimum span The manual's first action is: “Set the stop frequency to more than four times as many as the start frequency. And then select logarithmic sweep.”

Model
Agilent E5061A/E5062A ENA Series
Display shows
Log sweep requires 2 octave minimum span (53)
Type
Error code
Source
Manual p. 330

What 53 means

Log sweep requires 2 octave minimum span p. 330

Common causes

  • The span of sweep range is not satisfied the requirement for logarithmic sweep. p. 330

What to do

  1. Set the stop frequency to more than four times as many as the start frequency. And then select logarithmic sweep. p. 330

Before you service the product, read the safety instructions in the manual p. 9.

If 53 is still not resolved

  • Contact your Agilent Technologies Sales and Service Office for service and repair. p. 9
    “Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintain ed in operational condition.”

This page summarises what the manufacturer’s manual says about 53; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Agilent. Found a mistake? Report an error.