Thermo Fisher Scientific pH 6+ Manuals

ManualLib has 2 documents for the Thermo Fisher Scientific pH 6+ (Other Electronics): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
pH 6+
Documents
2
Languages
EN
Last added
2023-11-03

Start here

Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

User Manual 28 pages · 3.9 MB · EN

Thermo Fisher Scientific pH 5+, pH 6+, and Ion 6+ are microprocessor-based handheld meters with a large custom LCD. The pH 5+ measures pH and temperature in °C; the pH 6+ and Ion 6+ also measure mV (ORP) and temperature, and the Ion 6+ measures direct ion concentration of various mono- and divalent ions.

This manual explains keypad functions, LCD annunciators, battery replacement, electrode and temperature sensor connections, pH electrode conditioning, switching on, calibration, measurement, electrode care, troubleshooting, specifications, accessories, warranty, and return procedures.

Read online

Thermo Fisher Scientific pH 6+ specifications

Ion range0.01 to 1999 ppm
Ion resolution0.01 ppm for 0.01 to 0.99 ppm; 0.1 ppm for 1.0 to 199.9 ppm; 1 ppm for 200 to 1999 ppm
Ion accuracy+/- 1% of reading
Ion calibration points2 to 3 points (minimum 2 pts)
pH range0.00 to 14.00 pH
pH slope range80 to 120%
pH calibration points1 to 3 points (push-button)
Buffer optionspH 4.01, 7.00, 10.01 (USA); pH 4.01, 6.86, 9.18 (NIST); pH 4.10, 6.97 (Pb)
Temperature range0.0 to 100.0 °C
Temperature compensationAutomatic / Manual (0 to 100 °C)
Millivolt range-1000 to +1000 mV
Millivolt range-500 to 500 mV
Millivolt resolution0.1 mV for -200 to 200 mV; 1 mV for 200 to 500 mV
Millivolt accuracy+/- 0.2 and 2 mV respectively
Auto-buffer recognitionYes
Hold functionHO
Auto shut offAfter 17 minutes
Low battery indicationLO
DisplaySingle Custom LCD
Operating temperature0 to 50 °C
Power requirements4 x AAA alkaline batteries
Battery life500 hours
Meter dimensions/weight15.7 x 8.5 x 4.2 cm / 255 g

Source: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

All Thermo Fisher Scientific pH 6+ documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What does the LO annunciator indicate?

It indicates a low battery condition.

From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

How should a pH electrode be conditioned before use?

Soak it in pH 4 buffer solution for at least 1 hour, then rinse before use.

From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

What are the measurement modes for the pH 5+ and pH 6+ meters?

The pH 5+ measures pH and temperature (oC). The pH 6+ measures pH, mV (ORP), and temperature.

From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual

How do I calibrate the meter?

Calibrate using pH buffer standards (USA, NIST, or Pb groups) and follow the step-by-step procedure in Section 3.1. Calibration values are automatically saved.

From: Thermo Fisher Scientific pH 5+, pH 6+ and Ion 6+ Meters Instruction Manual